Table 3. Possible causes and correctiv e actions for ZWT problems
Symptom Possible Cause Corrective Actio n
USB cable not
securely connected
to ZWT
Remove and reconnect both
ends of USB Cable.
USB cable damag ed
Replace with Model 3141 USB
Cable only.
Green indicator light
on ZWT does not light
within 60 seconds of
powering on PRM
ZWT fault
Contact Boston Scienticusing
the information on the back
cover.
Telemetry RF signal
obstructed
Assure that a clear line-of-sight
path exists between ZWT
and pulse generator. Repeat
interrogation.
Telemetry RF signal
interference
Reposition or reorient ZWT at
least 7.6 cm (3 in) or further from
the PRM. Repeat interrogation.
USB cable not
securely connected
to ZWT and PRM
Remove and reconnect both
ends of USB cable. Repositio
n
wand over the pulse generato
r
and rep eat interrogation.
RF Telemetry fails
Reposition wand over
pulse generator and repeat
interrogation.
Telemetry: intermittent
communication
PRM software version
not current
Contact Boston Scienticusing
the information on the back
cover.
Reorient or re locate t he devices.
Increase the separation
between
the devices.
Connect the equipment t
oan
outlet on a different c
ircuit.
Tel em et r y:
interference
Harmful i nterference
caused by the ZWT or
the ZW T is negatively
impacted by other RF
devices
Contact Boston Scient
icusing
the information on the
back
cover.
UsinganExternalECGMonitorwiththePRM
Use the foll owing acc
essories to set up the conguration described in this
section:
Model 6750 Surface EC
G Patient Cable
Model 6629 ECG–BNC Slave Cable
Model 6577 Sterilizable Telemetry Wand